DS7008-POLYETHERSULFONE (PES) DIALYZER (HIGH FLUX DIALYZER)

SKU: DS7008 Category:
Add to Wishlist
Add to Wishlist

Description

Model and Specification DS7008-1 DS7008-2 DS7008-3 DS7008-4 DS7008-5 DS7008-6
Membrane Material Medical-grade polyether sulfone (PES)
Housing & Blood caps Material Medical-grade Polycarbonate (PC)
Potting Sealants Material Bi-component polyurethane sealants(PU)
Inner diameter of membrane (μm) 200
Wall thickness of membrane (μm) 40
Max. operation pressure (kPa/mmHg) 66.5/500
Max. flow rate (mL/min) Max. QB = 500mL/min, Max. QD = 500-800mL/min
Effective membrane area (m2) 1 1.2 1.4 1.6 1.8 2
UFR(mL/h*mmHg) 40 45 50 55 60 65
Blood compartment Volume (ml) 70 78 85 98 110 125
Clearance rate Test condition QB = 200mL/min
Urea 189 192 194 196 197 198
Creatinine 181 185 188 191 194 196
Phosphate 176 180 184 188 192 195
Vitamin B12 105 124 137 150 162 175
Inulin 65 78 94 107 120 130
Test condition QB = 300mL/min
Urea 248 255 262 270 276 280
Creatinine 225 235 243 250 256 262
Phosphate 224 230 236 242 248 254
Vitamin B12 128 142 156 170 183 196
Inulin 90 98 110 120 128 134
Test Condition QB = 400mL/min
Urea 297 306 315 324 330
Creatinine 274 282 290 297 304
Phosphate 260 270 280 290 298
Vitamin B12 154 170 185 198 212
Inulin 110 118 128 136 142
Sieving coefficient Test condition QB = 300mL/min,QF = 65mL/min
Albumin(%) ≤1
Inulin(%) 90±10
Myoglobin(%) 55±10
β2 -microglobulin(%) 70±10
Sterilization ɣ-rays

 

Note: Vitro performance parameter condition for UFR: Q=200mL/min, TMP=100mmHg, T=37°C, test by anticoagulation bovine plasma, total protein content 60 5g/l. Vitro performance parameter condition for clearance rate: Q=500mL/min Q=10mL/min T=37°C test by simulated body fluid (EN 1283) BDF