DS7009-Polyethersulfone (PES) Dialyzer (Low Flux Dialyzer)

SKU: DS7009 Category:
Add to Wishlist
Add to Wishlist

Description

DS7009-1 DS7009-2 DS7009-3 DS7009-4 DS7009-5 DS7009-6
Membrane Material Medical-grade polyether sulfone (PES)
Housing & Blood caps Material Medical-grade Polycarbonate (PC)
Potting Sealants Material Bi-component polyurethane sealants(PU)
Inner diameter of membrane (μm) 200
Wall thickness of membrane (μm) 40
Max. operation pressure (kPa/mmHg) 66.5/500
Max. flow rate (mL/min) Max. QB = 500mL/min, Max. QD = 500mL/min
Effective membrane area (m2) 1 1.2 1.4 1.6 1.8 2
UFR(mL/h*mmHg) 15 18 22 25 28 30
Blood compartment Volume (ml) 70 78 85 98 110 125
Clearance rate Test condition QB = 200mL/min
Urea 186 188 190 192 194 196
Creatinine 175 178 181 184 187 190
Phosphate 165 170 175 180 185 189
Vitamin B12 88 100 110 118 126 133
Test condition QB = 300mL/min
Urea 238 245 252 260 267 275
Creatinine 206 220 233 245 257 270
Phosphate 192 203 214 225 235 245
Vitamin B12 92 104 120 136 150 165
Test Condition QB = 400mL/min
Urea 268 279 290 300 312
Creatinine 248 261 275 288 300
Phosphate 230 244 257 270 283
Vitamin B12 110 125 142 158 173
Sterilization ɣ-rays

 

Note: Vitro performance parameter condition for UFR: Q=100mL/min, TMP=100mmHg, T=37°C, test by anticoagulation bovine plasma, total protein content 60 5g/l. Vitro performance parameter condition for clearance rate: Q=500mL/min Q=10mL/min T=37°C test by simulated body fluid (EN 1283) BD.